FACT¢â 5 with 3 M QZ |
|
|
|
Key Features: |
| 26 MHz - 40 GHz Performance Range |
| Full Compliance Testing for Radiated Immunity: - IEC 61000-4-3/EN61000-4-3 - ENV 50140, SAE J-1113 |
| Eligible for FCC 3 Meter Class B Facility Filing |
| Eligible for FCC 3 Meter Class B Facility Filing |
| Available as a Turnkey Package: Chamber, Tower, Turntable, Antennas, Instrumentation and Software |
ETS-Lindgren FACTTM 5 Chambers offer the ability to perform radiated emissions testing at a 5 meter distance. While not required by any current standard, measurements made at a 5 meter distance have better correlation to those obtained at a 10 meter distance. While permitting 5 meter testing, these chambers preserve the ability to perform full compliance testing of applicable 3 meter standards.
The chamber's small overall size results from the use of ETS-Rantec hybrid absorber ferrite tile, which require less volumetric space than traditional absorbers. ETS-Lindgren FACTTM 5 chambers will fit in many existing structures, yet are large enough to perform 4 meter antenna scans above the ground plane. When installed with ETS-EMCO LoProTM or EuroProTM turntables, pit excavations for motor assemblies are eliminated.
ETS-Lindgren FACTTM 5 chambers achieve their broadband performance using a unique arrangement of hybrif and ferrite tile absorber. The design was modeled using the proprietary numerical electromagnetic software responsible for creating chambers that set new standards for anechoic performance.
At 3 and 5 meter range lengths, ETS-Lindgren guarantees a Normalized Site Attenuation (NSA) better than ¡¾ 3.5 dB deviation from theoretical NSA within the quiet zone, over the frequency range of 30 MHz to 1 GHz. The quiet zone at 3 and 5 meter test distances is cylinder up to 2.0 meter in diameter, following the volumetric test procedure of ANSI C63.4-2000.
Anticipating tomorrow's requirements, the FerroSorbTM hybrid and FT series ferrite tile absorbers used in this chamber to also guarantee NSA performance of better than ¡¾ 3.5 dB deviation from theoretical NSA using the test method defined in ANSI C63.4 (2000) at the extended frequency range of 1 to 40 GHz.
Radiated Immunity Testing ETS-Lindgren's FACTTM 5 chambers can also be used to perform full compliance testing for IEC 61000-4-3/EN61000-4-3, ENV 50140 and SAE J-1113 requirements. At 3 meter range lengths, field uniformity of , + 6 dB is achieved in the test aperture over the frequency range of 26 MHz to 1 GHz. The test aperture is a vertical plane 1.5 meter x 1.5 meter at an elevation of .8 meter to 2.3 meters above the ground plane, following the field uniformity test procedure of IEC 61000-4-3.
Anticipating tomorrow's requirements, an optional absorber treatment package is available which permits radiated immunity testing at frequencies up to 40 GHz with , 6 + dB field uniformity, following the test procedure of IEC 61000-4-3.
Additionally, ETS-Rantec FerroSorbTM can safely withstand continuous field intensity of up to 200 V/m and intermittent field intensity of up to 500 V/m. This safely exceeds the field intensity requirements of most commercial RI tests.
Turnkey Systems ETS-Lindgren is an integrated manufacturer producing such well-known brands as ETS-Rantec absorber, Series 81 and ETS-Euroshield shielding and RF doors, ETS-EMCO antennas, towers, turntables, and test accessories. All of our products are designed to work together for maximum efficiency and performance.
|
|
Standard Configuration |
Eight corner-mounted light fixtures with floodlights (electrical distribution not included)
One single-leaf, Series 201, recessed contact mechanism (RCM), manually operated, RF-shielded door, 1.22 m x 2.13 m (4' x 7') clear opening
One door maintenance kit
Raised reflective ground plane, 60.96 cm (24")
Dielectric vapor barrier and masonite underlayment
3.18 mm (1/8") vinyl floor tile
Two access hatches in raised floor, 45.72 cm x 45.72 cm (18" x 18")
Six waveguide vents, 45.72 cm x 45.72 cm (18" x 18")
Three 2 x 30 amp, 60 Hz power-line filters for lights and EUT (wiring not included)
One connector panel, 15.24 cm x 91.44 cm (6" x 36") clear opening
Three "N" type connectors
One 3.81 cm (1.5") pipe penetration with flange nuts and cap
One fiber optic penetration kit
Design and fabricate RF-shielded enclosure, interior nominal shield-to-shield dimensions of 11.58 m L x 7.31 m W x 6.09 m H, (38' L x 24' W x 20' H) includes 30 cm (12") raised floor
One RF-shielded penetration for airline to MiniMastTM (compressed air is responsibility of customer)
Guaranteed performance and a five-year limited warranty; one-year warranty on doors, filters and moving parts; two-year warranty on optional ETS-Lindgren's EMCO brand equipment.
Shield test per MIL-STD-285, one frequency test at 1 GHz
ETS-Lindgren's Rantec brand FS-600 on the chamber ceiling and two side-walls
ETS-Ranted FS-400 on the two end-walls
Light reflective covers for FS-400 and FS-600 absorber
Light reflective covers for FS-400 and FS-600 absorber
Sixteen pieces of ETS-Lindgren's Rantec brand FT-100c ferrite tile panels for an area of the floor for radiated immunity testing up to 1 GHz
Sixteen pieces of ETS-Lindgren's Rantec brand EHP-18PCL pyramidal absorber on an area of the floor for radiated immunity testing from 1 GHz up to 40 GHz
Installation of the enclosure and absorber
Field uniformity calibration per IEC 61000-4-3 from 26 MHz to 1 GHz (optional test from 1 - 40 GHz
Normalized site attenuation test per ANSI C63.4-2000 from 30 MHz to 1 GHz (optional test from 1 to 40 GHz)
ETS-Lindgren's EMCO brand Model 2087-3.03 electrically powered, 3 m turntable with cable and fiber optic feed-through
ETS-Lindgren's EMCO brand Model 2090 dual device controller
ETS-Lindgren's EMCO brand Model 2075-2 MiniMastTM electrically powered air polarization boom tower with 20 m cable and fiber optic feed-through
One threaded brass ground stud 1.27 cm (0.5") diameter x 12.7 cm (5") long
|
|
Options |
Shielded control room, 3.05 m x 3.05m x 2.44 m (10' x 10' x 8')
Fire detection and suppression system
Electrical distribution
CCTV monitoring system
Seismic structural design calculations and certification
Radiated Immunity testing above 1 GHz
Intercom system
Preparation of FCC filing paperwork | |
Specifications |
ABSORBER TYPE |
INTERNAL WORKING AREA |
SHIELD ROOM INNER DIM |
OVERALL STRUCTURE DIM |
FS-600 FS-400 FT-100c ferrite panels EHP-18CL Absorber |
(L) 10.78 m .... 35.4 ft
(W) 6.12 m .... 20.0 ft
(H) 5.19 m .... 17.0 ft
|
(L) 11.58 m 38.0 ft
(W) 7.31 m 24.0 ft
(H) 6.09 m 20.0 ft
|
(L) 12.05 m 39.53 ft
(W) 7.78 m 25.53 ft
(H) 6.58 m 21.58 ft
|
FREQUENCY RANGE |
APPLICABLE STANDARDS |
IEC 1000-4-3 FIELD UNIFORMITY |
ANSI C63.4 NORMALIZED SITE ATTENUATION |
26 MHz to 40 GHz |
ANSI C63.4-2000 CISPR 11, 16, 22 EN50147-2 EN55016 EN55022 EN610000-4-3 FCC Class B parts 15 & 18 GR-1089-Core ISO 9000 MIL-STD 461E VCCI V3-/97.04 (1997) |
26MHz-18GHz 1.5 m x 1.5m +6dB 75% 16 points |
30 MHz - 40 GHz | | |
|
|